Implant Attachment
Instruments Used:
1. M6 Connection Screw (5)
2. 5 mm Hex Driver Shaft (25)
3. Blue-Handle Ratchet Driver (31)
Insert the M6 Connection Screw through the inner cylinder of the Targeting Frame (A). Attach the DynaNail
Implant onto the Targeting Frame, taking care to correctly align the keyed slots on the Implant with the
pegs on the Targeting Frame (B). Attach the 5 mm Hex Driver Shaft onto the Blue-Handle Ratchet Driver
(C) and use the Hex Driver to screw the M6 Connection Screw into the distal end of the DynaNail until
the Implant is tight on the Targeting Frame.