DynaNail® TTC Fusion System– Surgical Technique Guide DynaNail® TTC Fusion System– Surgical Technique Gu | Page 14

5 Guidewire Insertion Instruments Used: 1. Trocar Tipped Guidewire (19) 2. 3.3 mm Protection Sleeve (4) 3. 9.5 mm Protection Sleeve (3) 4. 14 mm Protection Sleeve (24) Assemble the 3.3 mm, 9.5 mm, and 14 mm Protection Sleeves as shown in Inset A. Place the Protection Sleeve Assembly against the plantar surface of the calcaneus. Insert the Trocar Tipped Guidewire (B) into the 3.3 mm Protection Sleeve, taking care that the guidewire tip is contacting the inferior aspect of the calcaneus. Drill the guidewire through the calcaneus and talus into the tibial medullary canal. With the Protection Sleeve abutted against the bone, confirm proper depth has been reached by reading the lasermarks off the back of the Protection Sleeve (C). When satisfied with the guidewire placement, remove the 3.3 mm Protection Sleeve leaving the Trocar Tipped Guidewire in place. TECH TIP • Fluoroscopy should be utilized throughout the entire insertion process to ensure centered guidewire placement in the tibial medullary canal on both the anterior and lateral views. A B 14 C