DynaNail® TTC Fusion System– Surgical Technique Guide DynaNail® TTC Fusion System– Surgical Technique Gu | Page 13

3 Nail Entry Site Preparation Make a plantar incision at least 2 cm in length. The incision can be in the longitudinal or transverse direction, according to surgeon preference. Bluntly dissect the soft tissue down to the surface of the calcaneus. Extreme care must be taken to protect the plantar neurovascular structures throughout the procedure. 4 Targeting Frame Assembly Instruments Used: 1. Targeting Frame (30) To assemble the Targeting Frame, slide the Manual Compression Tubes onto the Deployment Base. Push down on the gold latch on top of the Compression Tube arch to lock it in place (A). The Frame is secured in place when the Manual Compression Assembly no longer slides and the latch pops up to its original position. A BEFORE loading the Nail Implant, be sure to check that: • The numbered gold Stop Wheel is set to “0” and that the lever is in the upright position to allow the DynaNail Implant to attach to the frame (B). • The gold Manual Compression Knob is in the starting position by rotating it fully counter clockwise. This sets the Manual Compression Tubes to the “0” position (C). • If the Stop Wheel is not set to zero: pull the lever down, turn the Stop Wheel to “0,”and return the lever to its upright position. B C 13