PAGE 46
LT
Photonic Probe Station
version
vacuum shroud for photonic probe station incl. optical side access
PRODUCT KEY FEATURES
4
• Vacuum shroud for use with internal LT-APO low temperature
objective and lensed-fiber probes for side illumination & detection
of waveguide structures
• minimal sample drift, broadband high collection efficiency
• Sufficient space for 9 degrees of freedom to independently position
sample, focus, and 2 fiber probes
• Side access with additional window for RT optics
3
10
2
8
7
1
9
7
8
6
5
„ZOOM out“
Cryogenic Apochromatic Objectives
• negligible chromatic focal shift in working range
• uniform spot size and intensity within apochromatic range
...for further details, see page 128
1
2
3
4 attoDRY800 vacuum shroud for photonic probe station setup 6 ANPxy311/RES positioners for xy translation of sample and fiber probes
cold shield 7
8
9 2x ANPxyz51/RES for xyz translation of fiber probes
5 cold plate
top vacuum window 25 mm, cold shield top window 20 mm
additional window for side inspection 25 mm x 1 mm; window
material fused silica or N-BK7 (others on request)
10
LT-APO objective (see page 172 for details) on separate mounting post
ANPx311 for z focus of objective with thermal link for objective
feedthrough ring with blind flanges included, DC/SMA/fiber
feedthroughs optional